Preliminary
Findings on the Evaluation of Hand-Held Immunoassays for Bacillus
anthracis and Yersinia pestis
Counterterrorism
and Forensic Science Research Unit
Hazardous Materials Response Unit
Laboratory Division
Federal Bureau of Investigation
Quantico, Virginia
Bioterrorism
Preparedness and Response Program
Bioterrorism Rapid Response and Advanced Technology Laboratory
National Center for Infectious Diseases
Centers
for Disease Control and Prevention
Atlanta, Georgia
Abstract.......Introduction.......Sensitivity.......Specificity.......Repeatability.......
Robustness.......Stability.......Conclusions
Abstract
Commercially
available hand-held immunoassays (HHAs) for the detection of Bacillus
anthracis and Yersinia pestis (the causative agents of
anthrax and plague, respectively) were compared for sensitivity,
specificity, repeatability, robustness, and stability. A total of
six kits (four for anthrax and two for plague) were evaluated.
All
of the HHAs detected the Pasteur and Sterne strains of B. anthracis,
but an HHA from one manufacturer failed to detect a virulent strain
of B. anthracis. The lower limits of detection for the HHAs
varied from 105 to106 for unpurified anthrax
spores and from 105 to108 for washed anthrax
spores. The anthrax HHAs cross-reacted with nontarget bacteria,
yielding positive results in the absence of B. anthracis
from 3% to 83% of the time.
The
ability of the plague HHAs to detect Y. pestis was dependent
on the particular strain being analyzed as well as the temperature
at which the strain was grown. Each strain was grown at either 25°C
(77°F) or 37°C (98.6°F). No strain of Y. pestis
that was grown at 25°C was detected by the HHAs. The lower limit
of detection of the Y. pestis HHAs was between 104
and 105 cells. The plague HHAs cross-reacted with nontarget
bacteria, yielding positive results in the absence of Y. pestis
in 7% to 13% of the specimens tested. The results for sensitivity
and specificity demonstrate the limitations of these assays as a
reliable means of field detection.
Introduction
In
the interest of public safety, the following communication is being
released before subsequent full publication.
During
the emergency response phase of a suspected or potential exposure
to a biological threat, first responders sometimes use commercially
available hand-held immunoassays (HHAs) for the detection of biological
threat agents. HHAs are small test strips which contain antibodies
to a specific biological agent. A suspect sample is suspended in
a liquid supplied with the test assay. The liquid suspension is
then applied to the test strip and allowed to develop for approximately
15 minutes. If a biological threat agent is contained within the
sample, a colored band will appear on the test strip. A quality
control test is built into all the strips to indicate whether the
assay materials are working properly. The Federal Bureau of Investigation
and the Centers for Disease Control and Prevention tested six commercially
available HHAs (four for B. anthracis and two for Y. pestis)
in order to assess their capabilities and limitations under a variety
of anticipated scenarios. Test strips were used according to the
manufacturers' instructions. Effectiveness of commercially available
HHAs to detect B. anthracis or Y. pestis was evaluated
by measuring sensitivity, specificity, repeatability, robustness,
and stability. Preliminary results of this work are outlined in
this short communication. Out of consideration to the HHA manufacturers,
test strip names have been withheld. Names of test strips, complete
procedures, and full discussion of the results will be provided
in a detailed technical publication.
Sensitivity
In
order to establish the limit of sensitivity, the minimum number
of microorganisms required to give a positive test result is referred
to as the sensitivity of the assay. A stringent test was performed,
in which ten samples out of ten tested had to be positive. If fewer
than ten tests were positive at a certain concentration, then ten
additional samples were tested with ten times the previous amount.
This process was repeated until all ten samples tested gave positive
results.
For
anthrax sensitivity measurements, four different anthrax HHAs were
tested against unpurified B. anthracis spores. In addition,
a separate preparation of washed B. anthracis spores was
tested. Overall, the sensitivity values of the assays ranged from
105 to 108 spores of B. anthracis.
Two
plague HHAs were evaluated using 11 different strains grown at 25°
and 37°C. The Y. pestis HHAs only detected organisms
grown at 37°C because both HHAs are specific for a protein that
is produced only when the organisms are grown at temperatures greater
than 30°C. The sensitivity values of the assays ranged from
104 to 105 Y. pestis cells grown at
37°C. The lowest number of spores or cells required for visual
detection of a positive result for each strip is provided in Table
1.
Note
that the sensitivity values are much higher than those quantities
reported to produce disease (8,000 - 50,000 spores for inhalation
anthrax; 100 - 500 microorganisms for pneumonic plague).
See http:\\www.nbc-med.org\SiteContent\MedRef\OnlineRef\FieldManuals\
medman\Handbook.htm
Specificity
Specificity
refers to the ability of an assay to accurately discriminate between
a target microorganism (either Bacillus anthracis or Yersinia
pestis) and other bacterial species. In order to determine the
specificity, each assay was tested for cross-reactivity with a variety
of different bacterial species. Thirty non-B. anthracis species,
strains, and preparations were tested using the anthrax HHAs, and
15 non-Y. pestis samples were tested using the plague HHAs.
Under these conditions, any positive result observed would be considered
a false-positive result. Table 1 provides
the number of false-positive responses over the total number of
different tests performed. The results indicate a false-positive
rate between 3% and 83% for the anthrax assays and a false-positive
rate between 7% and 13% for the plague assays.
In
addition to these tests, both the anthrax and plague HHAs were tested
with different strains of the respective pathogen. Two virulent
strains and two vaccine strains of B. anthracis were tested
with the anthrax HHAs from three suppliers (numbers 1-3 in Table
1). All the anthrax assays tested except one gave positive results
with these samples. The HHAs of supplier 1 tested negative with
one virulent (disease-causing) strain. Eleven virulent strains of
Y. pestis grown at 37°C were tested. One HHA failed to
detect two of these strains, and another HHA failed to detect one
of these strains.
Repeatability
The
repeatability of an assay reflects its ability to produce the same
results under the same test conditions. For this analysis, all experiments
were performed in duplicate, and the repeatability of the results
is recorded in Table 1. The results
are reported as the number of times the tests were repeatable relative
to the total number of replicate tests performed. A high level of
repeatability demonstrates optimal performance. The repeatability
range for the HHAs tested was between 78% and 100%.
Robustness
An
assay that yields a very high number of correct responses when performed
under a variety of different experimental conditions is regarded
as robust. To determine the effect of interfering substances on
the HHAs, tests were performed both with and without target pathogen
in eight different solutions containing salt, sugar, milk, acid,
and alkaline solutions, and with two extracts of either Georgia
red clay or commercial potting soil. The ten tests were performed
in duplicate with each assay. Success rates for each solution are
shown in Table 1, listed as the number
of times that an expected result was obtained relative to the total
number of tests that were performed. There was no pronounced effect
of interfering substances on the performance of the new generation
HHAs. However, Georgia red clay gives a positive response with the
anthrax strips of suppliers 2 and 3 and reduces the intensity of
the reactions with the strip of supplier 1. High-salt solutions
caused the strips of supplier 3 to react inconsistently.
Stability
HHAs
were stored for 10 to 30 days at a field temperature of 30°C
(86°F) without losing sensitivity. These results indicate that
HHAs are generally stable. Therefore, special handling of test kits
is not required.
Conclusions
This
study was carried out to obtain information on the performance characteristics
of hand-held immunological assays for two pathogens, Bacillus
anthracis and Yersinia pestis. All HHAs tested gave false-positive
results with at least one bacterial species that was not the target
organism. These results indicate that further evaluation with environmental
samples collected from a wide geographical region is warranted to
better define the specificity of the HHAs.
The
number of organisms reported to produce disease in humans is much
fewer than the lower limit of consistent detection (105
to 108 spores of B. anthracis and 104
to 105 cells of Y. pestis) for both assays. Additionally,
some HHAs were incapable of detecting infectious agents that cause
anthrax and plague. For example, if highly infectious Y. pestis
in captured fleas (the host that carries the pathogen) were tested,
or organisms grown at room temperature (25°C), the results indicate
the HHAs would not react. Therefore, the HHAs cannot be relied upon
to always detect a pathogen when it is present, particularly the
plague assays.
The
HHAs generally performed well when substances known to interfere
with immunological tests were added or conditions such as pH (acidity)
and ionic strength (salt concentration) were changed. Results indicate
that guidelines for the testing of environmental samples are needed.
Variation in repeatability was observed with all anthrax HHAs, lowering
the level of confidence that can be placed in these assay systems.
The
low sensitivity, the high level of possible cross-reactivity with
environmental bacteria, and the failure to detect dangerous species
of B. anthracis and Y. pestis limit the usefulness
of these hand-held assays.
Top
of the page
|